BLOX. Here’s how BLOX helps protect your X-ray tube

September 2, 2025

X-ray tubes are precision components designed for years of service, but like any high-performance part, they have operating limits. One issue we sometimes see in the field is accelerated filament wear, especially in systems running at low kV and high tube current. Left unchecked, these operating conditions put heavy stress on the filament. While high current increases the number of X-rays produced, leading to a higher flux density and potentially faster acquisition times, it also puts more stress on the tungsten filament, shortening its lifespan. Conversely, a lower kV reduces the energy of the X-ray photons, potentially requiring a higher mAs value to achieve the same image quality, which can also impact tube life. 

With BLOX, you can see these patterns at a glance, adjust system use, and protect your hardware from premature failure.

However, when a system operates persistently to the left of the peak on the isowatt curve, low kV and high tube current, the filament is subjected to extreme thermal stress. This is particularly damaging when:

  • The system runs many exposure cycles per day in this range
  • The combination of kV and mA stays far from the optimal operating points

Over time, this results in excessive filament wear and eventually complete filament failure. Once the filament fails, the tube cannot be repaired—it must be replaced.

Indicates how frequently a set point is applied on the tube’s kV–mA characteristic chart.

How to spot it

BLOX makes it easy to detect high-risk operating patterns before they cause severe damage.

On the Exposure Details tab, you’ll find the isowatt curve alongside a record of the setpoints most commonly used by the system.

The larger the bubble, the longer the system spent at that setpoint. When exposures cluster heavily left of the curve peak—in the high filament wear zone—BLOX makes this pattern visible at a glance. This insight allows service teams to start a conversation with the end user about adjusting operating modes or selecting a tube better suited to the application.

Consequences

Excessive filament wear doesn’t just shorten tube life—it impacts system performance and customer operations:

  • Unplanned downtime occurs when the filament fails
  • Full tube replacement costs (repairs are not possible)
  • Decreased system availability during replacement and recalibration
  • Reduced customer satisfaction due to unpredictable failures

The financial and operational impact can be significant, especially for production-critical systems.

Recommendations

When BLOX indicates that a system is operating heavily in the high filament wear zone, the following steps are recommended:

  • Engage the end user to discuss their imaging requirements and explore if kV/mA setpoints can be adjusted without compromising results.
  • Evaluate tube options—in some cases, a different tube design is better suited to high-current operation.
  • Plan for contingencies—if no adjustments are possible, inform the end user about the risk of early failure and recommend keeping a spare tube in stock.

By bringing these patterns to light early, BLOX enables OEMs, service providers, and end users to collaborate on solutions that extend tube life, reduce costs, and ensure system uptime.

Protect your tube and your uptime

Excessive filament wear is avoidable, but only if you know it’s happening. BLOX gives you the clarity to see risky operating patterns before they cause expensive failures.

Don't wait for unplanned downtime to force your start using data to keep your systems healthy and your operations running smoothly. Talk to us today about how BLOX can help you spot problems early, extend tube life, and protect your hardware investment.

Latest three news

Spotlight on Shelton. Featuring Pierre Niepel

November 19, 2026

Meet Pierre Niepel, 20 years in medical imaging, now leading Comet X-ray sales in the Americas. Pierre Niepel has joined Comet X-ray as Head of Sales Americas, based at our U.S. site in Shelton, Connecticut. He spent about 20 years in the U.S. medical imaging market, working at Siemens Healthineers and Boston Imaging before running his own advisory business. Pierre brings experience from the equipment-manufacturer side of the industry, and a clear view of what customers expect from their suppliers. We asked him why he joined Comet, where he intends to start, and how he sees customer relationships developing.

Read more

Focus topic. Semiconductor segment insights

November 10, 2026

The semiconductor industry has mastered the art of precision. Nanometer-scale processes, extensive testing, and sophisticated control systems create an environment that appears highly predictable. But while process control continues to improve, many of today's most critical defects are becoming harder to see as they disappear inside complex, multi-layer structures. As advanced packaging reshapes the industry, visibility itself is emerging as a decisive factor in manufacturing success.

Read more

Welcome Andreea!

November 2, 2026

Meet Andreea Vasiliu, our new Vice President of Global Sales! Andreea brings extensive international sales and leadership experience and knows Comet well from her previous roles within Plasma Control Technologies. During more than eight years with the company, she held several global and regional sales leadership positions, most recently as Director Sales EMEA and Vacuum Capacitors Global. Since 2024, she has served as General Manager Europe and RoW at VAT Group.

Read more