Abrupt high-voltage shutdowns - here’s how BLOX helps prevent damage

September 23, 2025

High-voltage X-ray systems are built for performance and durability, but even robust hardware has its limits. One behavior we see far too often in the field is abrupt high-voltage shutdowns—often caused by well-meaning operators or misconfigured systems. This type of event may seem harmless in the moment, but over time it puts serious stress on the X-ray tube and can quietly reduce system longevity. With BLOX, we give OEMs, integrators, and end users the insight needed to detect these patterns early. The result is a smarter, more proactive approach to performance and uptime—driven by data, not guesswork.

Issue & cause

Our generators are designed to support the long-term health of the X-ray tube. Under normal operating conditions, the system initiates a controlled shutdown sequence where the high voltage ramps down gradually. This process ensures that the tube target cools in a stable, predictable way—minimizing mechanical and thermal stress.

But when high voltage is cut off abruptly, this controlled process doesn’t occur. The sudden stop creates an internal thermal shock that can quickly degrade the target material. Over time, this leads to premature wear, reduced image quality, and a shortened tube lifespan.

These irregular shutdowns are typically caused by three factors:

  • Activation of safety interlocks

  • Emergency shutdowns

  • Improperly configured customer interlocks

The root cause is often either human (such as an operator opening a door mid-exposure) or technical (like poorly timed PLC logic in the OEM’s system). Regardless of the trigger, the result is the same: unnecessary stress on the tube and lost value from the hardware.

Regular shutdown
Abrupt shutdown

How to spot it

One of the most useful features in BLOX is the alerts tab, which highlights abnormal behaviors such as abrupt high-voltage shutdowns. In the case of one customer installation, BLOX logged over 2,000 events tied to the safety interlock. Each time, the door was opened during exposure, cutting power too quickly and skipping the protective ramp-down.

The system doesn’t just flag the issue—it explains the nature of the event clearly. These alert messages are customized to call attention to lifetime-reducing behaviors so that OEMs and service teams can prioritize the right corrective actions. Over time, fewer alerts mean better uptime and more predictable maintenance planning.

Screenshot of the alert tab

Irregular shutdowns

To go further, BLOX provides a daily events chart that makes irregular shutdowns easy to spot and analyze. This graph offers a quick overview of system behavior with intuitive visual markers:

  • Green bars represent normal, controlled shutdowns

  • Yellow bars indicate irregular shutdowns

By comparing these patterns, you can quickly assess how often and when irregular shutdowns occur. Hovering over each yellow bar reveals the specific reason for that day’s anomaly—whether it was a safety interlock, temperature issue, or another trigger.

This kind of visibility is crucial. It gives OEMs and service teams the data they need to have informed conversations with end users and to fine-tune system behavior before any real damage occurs.

Screenshot of the daily events graph

Consequences

While a single abrupt shutdown may not cause visible harm, repeated events create a cumulative effect. Each time the system shuts down without properly ramping down high voltage, the X-ray tube is exposed to additional thermal stress. Over time, this accelerates wear on the target and degrades the focal spot.

The result is often:

  • Dose loss during imaging

  • A drop in image quality

  • Increased exposure time requirements

  • Ultimately, a reduction in overall system performance

BLOX helps prevent this by making the invisible visible. By catching irregular shutdown patterns early, BLOX allows OEMs and users to act before these small events turn into costly service calls or premature component replacements.

Before thermal stress
After thermal stress
Before thermal stress
Gradual dose loss caused by thermal stress
Advanced dose loss caused by thermal stress

Recommendations

When BLOX indicates a recurring pattern of irregular shutdowns due to safety interlocks, the next step is to investigate the root cause. In most cases, the source will fall into one of two categories, often due to safety interlocks:

  • A human factor, such as an operator interrupting exposures without realizing the consequences

  • A technical issue, such as poorly tuned PLC logic, door sensor misalignment, or unintentional timing conflicts during system integration

Once the cause is known, the response can be simple. Often, it means retraining personnel, improving interlock logic, or adjusting exposure workflows. In some cases, a service visit may be required to correct system-level behavior.

Whatever the solution, the key is clarity—and that’s exactly what BLOX delivers. By highlighting irregular shutdowns clearly and early, BLOX supports smarter service decisions, stronger OEM and end-user collaboration, and better protection of your hardware investment.

Latest three news

Welcome Andreea!

November 2, 2026

Meet Andreea Vasiliu, our new Vice President of Global Sales! Andreea brings extensive international sales and leadership experience and knows Comet well from her previous roles within Plasma Control Technologies. During more than eight years with the company, she held several global and regional sales leadership positions, most recently as Director Sales EMEA and Vacuum Capacitors Global. Since 2024, she has served as General Manager Europe and RoW at VAT Group.

Read more

Need for Speed: Boosting X-ray Inspection Speed in Advanced Electronics

October 28, 2026

As electronics become more powerful and advanced packaging technologies more complex, manufacturers face the challenge of detecting ever-smaller defects without reducing throughput. Optical inspection methods cannot reveal the buried structures found in high-power electronics, chiplets, HBM, through glass vias (TGVs), and advanced 3D packages. X-ray inspection and automated X-ray inspection (AXI) provide the required visibility, but higher resolution and penetration have traditionally come at the expense of speed. Recent advances in micro and nano focus X-ray source technology are overcoming this trade-off by delivering higher photon flux, improved contrast-to-noise ratio, and faster imaging. In this post we review these developments in high-power electronics, where higher voltage and power are required to penetrate through dense structures while maintaining micro-scale imaging, and advanced packaging where increasing complexity from vertical integration and sub-micron resolution combine. We highlight how increased throughput can be achieved while maintaining reliable defect detection.

Read more

Focus topic. A conversation with Jan Bressendorff and Allen Gress about pipeline weld inspection

October 8, 2026

On a pipeline job, the X-ray system is chosen for the weld, not from a catalog. Pipe size and access determine how the crew can inspect it. The procedure sets what the image must show. Heat, weight and exposure time shape the working day. A poor choice can cost hours. A good one helps the crew get an image the inspector can assess without needless delay.

Read more